Field safety notice FSN-UK-2021-03
VR-5 IV pole socket insert
- Issued
- Products
- VR-5
- Status
- Closed
- Last updated
Products affected
VR-5 beds with corner socket inserts from batch 2019-44 to 2020-31.
What has happened
A batch of IV pole socket inserts was moulded from a material with lower impact strength than specified. If an IV pole carrying more than its rated load is knocked, the insert can crack and the pole can lean.
Action to be taken by the user
- Do not load IV poles on affected beds beyond 12 kg.
- Your Van Roon engineer will replace the inserts at the next visit, or sooner on request.
Contact
For questions about this notice, contact the service desk on 0808 157 0614 (Monday to Friday, 08:00 to 17:30) or service@vanroon.uk, quoting FSN-UK-2021-03.
The competent authorities concerned have been told about this notice. This notice has been sent to every hospital on our records with the products affected.